Tesis

Perbandingan Nilai Amplitudo dan Latensi Antara Elektroda Jet, Elektroda Dencott dan Elektroda Dawson-Trick-Litzkow Pada Pemeriksaan Elektroretinografi Multifokal Mata Normal Pada Populasi Orang Dewasa di Rumah Sakit Umum Pusat Nasional Cipto Mangunkusumo = Comparison of Amplitude and Implicit Time Between Jet Electrode, Dencott Electrode and Dawson-Trick-Litzkow Electrode in Multifocal Electroretinography Examination in Adult in Cipto Mangunkusumo National Hospital.

Mendapatkan nilai standar normal pemeriksaan elektroretinografi multifokal dan mencari perbandingan nilai tersebut berikut tingkat kenyamanan pasien menggunakan elektroda Jet, Dencott dan Dawson-Trick-Litzkow pada populasi orang Indonesia dewasa di RSUPNCM. Melalui metode convenient sampling dari 49 subjek orang Indonesia berusia 19-49 tahun dengan kedua mata normal, dilakukan pengukuran amplitudo dan latensi gelombang berdasarkan rekomendasi ISCEV. Pengukuran meliputi gelombang N1 dan P1 pada ring 1 sampai 5. Terdapat perbedaan signifikan nilai standar normal pemeriksaan elektroretinografi multifokal pada populasi orang Indonesia dewasa secara statistik antara elektroda dengan amplitudo gelombang lebih tinggi dan latensi lebih panjang pada elektroda Jet dan Dencott dibandingkan dengan DTL. Elektroda DTL dan Jet dirasakan lebih nyaman daripada elektroda Dencott bagi orang Indonesia dewasa normal untuk pemeriksaan elektroretinografi multifokal. Elektroda DTL memberikan amplitudo gelombang yang paling rendah, latensi yang apling pendek dan tingkat kenyamanan yang paling tinggi dibandingkan elektroda Jet dan Dencott dalam pemeriksaan elektroretinografi multifokal pada orang Indonesia dewasa normal.
Kata kunci: Elektroretinografi multifokal; elektroda Jet; Dencott; Dawson-Trick-Litzkow; tingkat kenyamanan; Indonesia dewasa normal


To establish normal values of multifocal electroretinography (ERG) and to compare the values and the comfort level using Jet, Dencott and DTL electrode in Indonesian Adult. Through convenient sampling 49 normal Indonesian subjects between 19 and 49 yeras oldwere selected. Multifocal ERG amplitudes and implicit time values were measured according to recommendation by the International Society for Clinical Electrophysiology of Vision. Evaluation consisted of N1 and P1 wave in ring 1 to 5. after the examination, all subjects filled in a questionnaire about comfort level, adopted from the visual analog scale. We observed a statistically significant difference in multifocal ERG normal values between electrode with higher wave amplitudes and longer implicit time in Jet and Dencott electrodes compared to DTL electrodes. Jet and DTL electrodes are more comfortable than Dencott electrodes for Indonesian adults in multifocal ERG. DTL electrodes give the lowest wave amplitude and the shortest implicit time and are the most comfortable electrode compared to Jet and Dencott electrodes, in multifocal ERG in Indonesian adults.
Keywords : Comfort level; Dawson-Trick-Litzkow; Dencott; Jet electrode; Multifocal electroretinography; normal Indonesian adult.

Judul Seri
-
Tahun Terbit
2018
Pengarang

Ika Citra susanti - Nama Orang
Mohamad Sidik - Nama Orang
Syntia Nusanti - Nama Orang
Aria Kekalih - Nama Orang

No. Panggil
T18554fk
Penerbit
Jakarta : Program Studi Ilmu Kesehatan Mata.,
Deskripsi Fisik
xiii, 68 hal; ill; 21 x 30 cm
Bahasa
Indonesia
ISBN/ISSN
-
Klasifikasi
T18554fk
Edisi
-
Subjek
Info Detail Spesifik
-
T18554fkT18554fkPerpustakaan FKUITersedia
Image of Perbandingan Nilai Amplitudo dan Latensi Antara Elektroda Jet, Elektroda Dencott dan Elektroda Dawson-Trick-Litzkow Pada Pemeriksaan Elektroretinografi Multifokal Mata Normal Pada Populasi Orang Dewasa di Rumah Sakit Umum Pusat Nasional Cipto Mangunkusumo = Comparison of Amplitude and Implicit Time Between Jet Electrode, Dencott Electrode and Dawson-Trick-Litzkow Electrode in Multifocal Electroretinography Examination in Adult in Cipto Mangunkusumo National Hospital.

Related Collection